Influence of the scleral lens and fluid reservoir thicknesses on residual astigmatism


Article
Version publiée / Version of Record
Loading...
Thumbnail Image

Contributor(s)

Advisor(s)

Published in

Journal of contact lens research and science

Conference Date

Conference Place

Publisher

Dougmar Publishing Group

Degree Level

Discipline

Keywords

  • Scleral lenses
  • Residual astigmatism
  • Lens flexure
  • High-order aberrations

Funding organization(s)

Abstract

Purpose This study aims to determine if lens or tear fluid reservoir thicknesses (LT/FRT) may influence the presence of residual astigmatism and participant’s visual acuity. Methods The study was a randomized, non-dispensing, prospective study. Empirically and randomly chosen participants were fitted with 4 combinations (350 and 250 um LT fitted with 250 and 350 um FRT) of 16 mm diameter scleral lenses, designed using a corneo-scleral profiler software (sMap 3D, Visionary Optics, US). Lenses haptics were kept spherical for all lenses. They were evaluated under a slit lamp, anterior segement OCT (objective fluid reservoir and lens thicknesses), topography over lenses and aberrometry, after 30 minutes of lens wear. Spherico-cylindrical refraction and logMar acuity were also assessed. Results Study population was composed of 24 participants aged 24.2 + 4.7 years old. Baseline refractive error was -2.3 + 1.6 D with -0.48 + 0.26 D of astigmatism. In vivo (OCT) lens A was 344.1 ± 15.4 um thick, fitted with a vault of 213.6 ± 42.4 um; Lens B was 346.2 ± 12.5/327.2 ± 44.8; Lens C was 260.3 ± 17.7/214.0 ± 40.6 um and Lens D was 262.2 ±13.2/330.8 ± 52.0 respectively. All lenses were found similarly decentered inferiorly by 0.10 to 0.15 um. BCVA was −0.32 + 0.08 (A), −0.21 + 0.10 (B), −0.28 + 0.08 (C), and −0.14

  • 0.10 (D), compared to −0.25 + 0.08 (A), −0.11 + 0.10 (B), −0.23 + 0.06 (C), and −0.05 + 0.12 (D) when sphere only was compensated. Residual refractive astigmatism (RA = -0.50 to -0.75D) is found significantly higher based on the FRT (F=9.560; p=0.037) and not LT(F=0.429; p=0.522). There is no correlation between RA and over-k readings (Lens A r=-0.078, p=0.773; Lens B r=−0.073, p=0.788; Lens C r=−0.345, p=0.171; Lend D r=0.019, p=0.944). Higher order aberrations, mostly vertical coma, were found clinically significant but not statistically different between lenses (A= 0.350 + 0.032; B=0.382 + 0.053, C=0.329 + 0.044 and D=0.385+ 0.062; p=0.776) Conclusion This study proves that low level of RA may be found when scleral lenses are fitted on normal corneas. Its occurrence is related to the presence of high-order aberrations and less likely to lens flexure.

Table of contents

Notes

Notes

Other language versions

Related research dataset(s)

Endorsement

Review

Supplemented By

Referenced By

This document disseminated on Papyrus is the exclusive property of the copyright holders and is protected by the Copyright Act (R.S.C. 1985, c. C-42). Unless the document is published under a Creative Commons licence, it may be used for fair dealing and non-commercial purposes, for private study or research, criticism and review as provided by law. For any other use, written authorization from the copyright holders is required.